LOGIC TEST (See Fig. 8)
The logic test indicates a TTL high, low or changing logic signal. Turn the
range switch to the logic position. The LCD will display a I0L and the word
“Logic.” Connect the black test lead to circuit common or ground point and
the red lead to the test point. A logic LOW will display the down arrow a with a
tone, a logic HIGH the up arrow without tone, and a changing signal will alternate
the arrows and tone. Above 15Hz the tone may be intermittent then go steady
above 100Hz. Above 200Hz the up/down arrows both stay on solid. See the “
MEM” button instructions for detecting fast pulses and waiting for one-shot
changes without having to observe the meter.