User Manual 301/302
trigger advances to next programmed sequence.
Control input is TRIG IN connector.
Advance Source
Internal Internal programmable rate generator
Period From 20 µs to 1000 s
Accuracy ±(1% + 1 µs)
External
Input Front panel Trigger input
Frequency 15 MHz to DC
Software IEEE 488.2 command
OUTPUTS (CHANNELS A & B)
Connector Front panel BNC
Stand-by Output Off or Normal
Impedance 50Ω, ±1%
Protection Protected against temporary short to case ground
Glitch Energy 100 pV-s at 5 Vp-p
Amplitude
Range 10 mV to 5 Vp-p, into 50Ω
Resolution 3.5 digits
Accuracy (1 KHz) ±(1% + 25 mV), 1.000 V to 5 Vp-p
±(1% + 5 mV), 100 mV to 999.9 mVp-p
±(1% + 2 mV), 10 mV to 99.99 mVp-p
Offset
Range 0 to ±2.495 V
Resolution 5 mV
Accuracy ±(2%+10 mV)
Filters
150 MHz, Elliptic 7 pole - standard
70 MHz, Elliptic 7 pole - standard
5 MHz - available upon request
2.5 MHz - available upon request
800 KHz - available upon request
Square Wave, Pulse
Rise/Fall time <2.5 ns, 10% to 90% of amplitude
Aberration <5%
Specifications