
Reverse Path Troubleshooting: CPD (Common Path Distortion) Analysis
101
5
Technicians need to be able to associate seemingly “unconnected”
symptoms. For example, someone notices a minor amount of hum in the
picture of a customer; two blocks away error rates start going up on a data
modem or DMX box. Somewhere in the area you find a connector cracked at
the housing entry which has allowed moisture entry and oxidation to occur.
Note that this can and does happen to F connectors in homes or institutions.
To solve such problems, the source of the moisture ingress needs to be
found and removed, the moisture removed and the oxidation or corrosion
removed. It’s likely to be cheaper, easier and faster to remove and replace
an entire device (tap, directional coupler or amplifier) than to determine and
fix the cause of the moisture ingress. If a bad F connector is found, it is good
practice to also replace the device to which the F connector was attached. In
the case of F connectors, relocation of equipment to be out of the path of
water and/or use of higher cost F connectors with better water seals is
recommended (it’s silly to save 25 cents on an F connector and generate a
50 dollar trouble call). (David Devereaux-Weber, University of Wisconsin -
Madison, (djdevere@facstaff.wisc.edu))
Recommended Setup for CPD Analysis
If the test point is not band limited to the reverse spectrum (in other words,
on the low band side of a diplex filter) a low pass filter may be needed to
eliminate the possibility of instrument related intermodulation products that
mimic CPD. The susceptibility to intermod is dependent on the total number
of carriers present on the input to the instrument. The measurement
technique is similar to using a band pass filter for intermodulation tests
(composite second/third order).
See Also
½ "CSO/CTB Measurements" in Chapter 6 explains how to
make composite second/third order measurements
(page 128).
The objective is to avoid bombarding the RF input of the instrument with a
large number of relatively high level signals (typically 78 channels in a 550
MHz system), while looking for low level signals. The problem is that when
attenuation is removed to see the low level beat products more clearly, the
front end of the instrument may be overloaded with the high level signals on
the same test point, and generate beat products of its own.
You might need a low pass filter when you're connecting the instrument to a
bi-directional amplifier test point—a test point with both forward and reverse
signals present. This phenomenon may present itself when performing a
reverse noise test using the SDA-5500 transmitter. To minimize the
Komentarze do niniejszej Instrukcji