Description and Basic Operation of the SDA-5500 and 5510: Basic Modes of Operation
246
13
Fig. 13-21 The Operator Information Screen
Frequency Adjustments
The FRQ soft key is used to set the start and stop frequency (for display
purposes only). A Start and Stop soft key is displayed to select the
parameter to be changed. You can be increment or decrement the frequency
with the diamond keys, or enter a specific value using the numeric keys
followed by the Enter key.
From the FREQUENCY submenu, you can use the Marker Zoom function to
change the start and stop frequencies so that they coincide with the marker
value. Use the Undo soft key to return to the original start and stop values.
Press the FRQ soft key again to return to the main sweep display.
Level Adjustments
The LVL soft key is used to adjust the vertical parameters of the graph.
These parameters include the reference level and the scale. The reference
level is the midpoint level displayed on the graph. It can be adjusted using
the diamond keys or by entering a numeric value followed by the Enter key.
The scale parameter (1,2,5, & 10 dB/div) can be adjusted only with the
cursor keys. For example, if the reference level was set at 2 dB and the
scale was set at 5 dB/div, the first horizontal grid line above the center would
be equal to 7 dB. Use the Auto Scale soft key to set the reference level
automatically for an optimum sweep display. Pressing the FCN and Scale
keys will also automatically scale the sweep display. Press the LVL soft key
to return to the main sweep display.