Wavetek 147 Dokumentacja Strona 325

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 324
SDA-5500/SDA-5510 Sweep Configuration: Sweep History and Theory
301
14
Another difference between forward and reverse sweep alignment is that
signals in a reverse path plant are TDMA digital channels: multiple
transmitters are sending to one receiver at the headend. For these systems
to work, the reverse carriers must be turned off when not in use. Accordingly,
the sweep methodology that allows accurate forward path sweep without
injecting any signal cannot be used here. Sweep pulses must be injected.
To prevent any chance of system interference, Acterna recommends that
you not put any sweep pulses within active return path channels. However,
due to the narrow frequency pulse used by the Stealth system, the edges of
channels can almost always be used without creating any interference
problem. For example, if two return path telephony carriers are located at
21.0 to 21.6MHz, and 21.6 to 22.2MHz, sweep points can be injected
between and around the two channels at 21MHz, 21.6MHz, and 22.2MHz.
Because cable systems are designed with appropriate spacing for the
forward high frequency range, the reverse path may not require amplification
at each station. In the cable, lower frequencies are attenuated less than
higher frequencies.
Productivity Tools
Stealth sweep incorporates several productivity tools, described below, that
help you work more effectively.
Markers and Limits
Stealth sweep systems incorporate several Marker and Limit functions.
Vertical markers: Two vertical markers enable you to check frequency
and level at specified points on the sweep. The display also shows the
difference values. These vertical markers define the minimum and
maximum frequency range to measure.
Horizontal markers: Two horizontal markers can be turned on or off.
When they are enabled, the horizontal markers show the maximum and
minimum response values on the sweep curve between the vertical
markers. Use these to automate maximum/minimum testing to find a
systems flatness.
Sweep limits: You can set a limit value for the largest deviation from
flat, and the meter will automatically display whether the current test
point passes or fails. Activate this by selecting the Limit Check mode
within sweep, and then setting the Limit Value (in dB).
Sweep Limit Variable: A common formula for system flatness is N/
10+X. This means that the response at the Nth amplifier in the system
Przeglądanie stron 324
1 2 ... 320 321 322 323 324 325 326 327 328 329 330 ... 409 410

Komentarze do niniejszej Instrukcji

Brak uwag